Customization: | Available |
---|---|
After-sales Service: | 24 Global Services |
Power Supply: | 220V |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product Name | Constant temperature and humidity weighing system |
Product Model | KS-FTHW408L |
Quantity | 1 set |
Weight | 450KG (approx.) |
Applications | A high precision balance is placed inside a high precision constant temperature and humidity system, the sample to be weighed is placed inside the system, balanced for 24 hours and then weighed fully automatically. This system overcomes the effects of changes in temperature and humidity in the laboratory environment on the weighing results and improves the accuracy of the weighing. |
Features | The constant temperature and humidity weighing system case comes with its own pulley and self-locking function, which makes it easy to move and fix the whole machine, and the balance works within the system. The sample is balanced in a high precision constant temperature and humidity environment to ensure the accuracy of sample weighing. The system has an automatic alarm function for mechanical faults and an automatic alarm function for lack of water in the water tank. The system is reasonably designed and uses multi-stage anti-vibration measures to ensure that the balance works continuously and smoothly and the readings are stable. The humidifier adopts ultrasonic humidification mode, closed circuit circulation, the condensate generated by dehumidification flows back to the same water supply tank to solve the problem of frequent water refilling. The wind speed inside the chamber does not affect the storage of samples, taking, etc. The constant temperature and humidity adopts the cyclic mode of air supply and suction to ensure uniform temperature and humidity inside the chamber. The observation window is made of double-layer hollow toughened glass, which is transparent, heat-insulating and not easy to produce steam frosting, etc.; silicone gloves are configured to ensure that the temperature and humidity fluctuations in the chamber are in accordance with the standard requirements of HJ836-2017 Determination of low concentration particulate matter in stationary pollution sources by weight during manual operation, and the weighing of the balance is not affected. |
Basic working principle | 1. Heating principle: the heating in the test chamber is mainly the use of high-power PPT pulse heating body, to improve the heating and heating precision, heating more balanced. 2. Refrigeration principle: the refrigeration method of the test chamber is mainly mechanical refrigeration, mechanical refrigeration using compression refrigeration, they are mainly composed of imported compressors, condensers, evaporators, etc. 3. Humidification principle: built-in ultrasonic humidifier humidification atomization. |
Dimensions | Inner box size: 1445×725×1700mm (L*W*H) Refrigeration unit: 500×500×1200mm |
Temperature control range | 15ºC~70ºC Arbitrary adjustment (factory setting is 25ºC) |
Temperature control accuracy | ±0.1ºC |
Temperature resolution | 0.1ºC |
Humidity range | 25% to 80% R.H (50±5% R.H relative humidity) (factory setting 50% R.H) |
Temperature control accuracy | ±0.1ºC |
Temperature resolution | 0.1ºC |
Humidity range | 25% to 80% R.H (50±5% R.H relative humidity) (factory setting 50% R.H) |
Humidity fluctuation | ±2% R.H |
Humidity resolution | 0.1% R.H |
Average rate of temperature rise and fall | 0.7ºC~1.0ºC/min |
Total installed power | 2.8kw |
Power requirements | 220V±10%; 50 Hz |
Constant temperature and humidity chamber | 1 set |
Double air duct constant temperature and humidity system (built-in) | 1 set |
Refrigeration unit (external) | 1 set |
Filter membrane placement tower | 2 units |
Anti-vibration sheet metal table | 1 unit |
Ion purging and de-staticising device (built-in) | 1 set |
Industrial control computer and workstation | |
Laser printer | |
Fully automatic mechanical operating arm |